4 Maintenance Procedures
138
March 14, 2014, 715004387 Rev. B
2. Remove the ion block assembly from the PEEK ion block support (see
3. Use the 3-mm Allen wrench to unscrew and remove the 4 screws
securing the PEEK ion block support to the adaptor housing, and
remove the ion block support.
4. If any of the O-rings show signs of deterioration or damage, dispose of
them in accordance with local environmental regulations.
Warning:
To avoid spreading contamination, dispose of the
O-rings in accordance with local environmental regulations.
The O-rings can be contaminated with biohazards or toxic
materials.
Securing screws
Adaptor housing
PEEK ion block
support
Ion guide
assembly