Connecting the liquid waste line
March 14, 2014, 715004387 Rev. B
233
Connecting the liquid waste line
Required material
Chemical-resistant, powder-free gloves
To connect the liquid waste line:
1. Place a suitable waste container below the instrument.
2. Slide a drain line over the barbed fitting of the drain (located at the
bottom of the instrument).
Warning:
To prevent personal contamination with biohazards or
toxic materials, always wear chemical-resistant, powder-free
gloves while performing this procedure. The waste line and
connection can be contaminated.
Notice:
To avoid distorting the drip tray or causing the drain cup
to leak, restrain the drain cup when attaching or removing the
waste line.
TP02684
Drain cup
Barbed fitting
Drain line