D Plumbing the IntelliStart Fluidics System
250
March 14, 2014, 715004387 Rev. B
Tubing and connection specifications
When replacing tubing and connectors, refer to the tables below for the correct
specifications.
Important:
The tubing for the sample reservoirs (ports A and B) is not
user-serviceable. To replace the tubing, contact Waters to arrange an engineer
visit.
Notice:
To avoid leaks, when plumbing the IntelliStart Fluidics system,
use only the tubes, nuts, and ferrules shown in the following tables, to
ensure over-pressure protection. The use of parts not recommended here
can result in leaks that the built-in leak sensor does not always detect.
Replacement tubing specifications:
Port
ID
(inches)
Color
Length (m)
R, A, and B
0.020
Orange
0.65
W
0.005
Red
0.2
S
0.004
Black
0.4
LC
0.004
Black
As required
Replacement nut and ferrule specifications:
Port
Nut code
Ferrule code
P
P245
P200
LC
F196
F192
S, W, R, A and B
F130
-