NanoFlow source
March 14, 2014, 715004387 Rev. B
79
Recommendation:
When using a GELoader tip, break the glass
capillary in half, scoring it with a fused silica cutter so that the
GELoader can reach the capillary’s tip.
9. Thread the knurled nut and approximately 5 mm of conductive
elastomer over the blunt end of the capillary.
10. Fit the capillary into the holder (probe).
11. Finger-tighten the nut so that 5 mm of glass capillary protrude from its
end.
Tip:
Measure the protrusion from the end of the nut to the shoulder of
the glass capillary.
Sprayer assembly:
12. Screw the sprayer back into the assembly.
13. Replace the sprayer cover.
14. On the ES+/- Source tab of the MassLynx MS Tune window, ensure that
the Capillary parameter is set to 0 kV.
5 mm
PTFE tubing
Ferrule
Union
Knurled nut
Blue conductive elastomer
Glass capillary