APCI mode
March 14, 2014, 715004387 Rev. B
65
Removing the IonSABRE II probe
Required material
Chemical-resistant, powder-free gloves
To remove the IonSABRE II probe:
1. Prepare the instrument for working on the source (see
).
2. Remove the corona pin (see
).
3. Disconnect the selector valve tubing from the IonSABRE II probe.
4. Unscrew the probe locking ring.
5. Carefully remove the probe from the probe adjuster assembly.
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while performing this procedure. The LC
system connections, IonSABRE II probe, and source can be
contaminated.
Warning:
To avoid electric shock, before beginning this procedure,
prepare the instrument according to the procedure on
,
“Preparing the instrument for operations on or inside its source”
.