4 Maintenance Procedures
88
March 14, 2014, 715004387 Rev. B
Preparing the instrument for operations on or inside its
source
For safety reasons, you must follow this procedure before working on the
source (for example, when changing the probe, operating the source isolation
valve, and maintaining the source).
Follow the procedure for the software that controls your mass spectrometer:
MassLynx software, or UNIFI software.
To use MassLynx software to prepare the instrument for operations on or
inside its source:
1. In the Instrument Console, click Stop Flow
to stop the LC flow or, if
column flow is required, divert the LC flow to waste as follows:
a. In the Instrument Console system tree, expand Xevo TQD,
Interactive Fluidics.
b. Click Control
.
c.
Select Waste as the flow state.
2. In the Instrument Console, click Standby
, and confirm that the
Operate indicator is not illuminated.
3. Wait 3 minutes to allow the desolvation gas flow to cool the probe and
source.
4. In the Instrument Console, click API
to stop the desolvation gas
flow.
5. Lift the visor on the front of the instrument so that it is clear of all the
source components and probe.
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...