4 Maintenance Procedures
166
March 14, 2014, 715004387 Rev. B
Cleaning or replacing the corona pin
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
HPLC-grade (or better) methanol
•
Lint-free tissue
•
Lapping film
•
Corona pin
To clean or replace the corona pin:
1. Remove the corona pin from the source (see
for damage.
2. Replace the corona pin if it is damaged; otherwise clean its tip with the
lapping film and a methanol-saturated tissue.
3. Install the corona pin in the source (see
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the probe and source
components.
Warning:
To avoid burn injuries, take great care while performing
this procedure; the probe and source can be hot.
Warning:
To avoid electric shock, ensure that the instrument is in
Standby mode before commencing this procedure.
Warning:
To avoid puncture wounds, handle the corona pin with
care; the corona pin tip is sharp.
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...