Replacing the column-outlet tubing in a CM-30S installed in a UPC
2
system
October 7, 2014, 715004178 Rev. B
93
4. If you are not installing a column on the column-outlet tubing that you
are replacing, route the tubing through a column clip on the column
plate.
Rationale:
Doing so prevents loose tubing from interfering with the
operation of the column-compartment drawer and outer CM-30S door.
5. If you are installing a column on the column-outlet tubing that you are
replacing, do as follows:
a. Ensure the cap nut is loose on the column fitting, insert the
column-outlet tubing into the column, and then tighten the column
fitting.
Column-outlet tubing
Column plate
Column clip
Cap nut
Column fitting
Column
Column-outlet tubing
Summary of Contents for ACQUITY CM-30S