3 Maintaining the 30-cm single-zone column manager
92
October 7, 2014, 715004178 Rev. B
3. Bend the tubing so that it immediately extends upward after the fitting,
and then back towards the rear of the drip tray, as shown in the figure
below.
Note:
The bending ensures that this section of tubing is the lowest
portion exiting from the valve so that any drips fall into the drip tray.
Correct bending and routing of tubing from outlet valve:
Important:
•
Route tubing from ports 1, 7 (W), and 6 up and over the outlet valve
and then down through the opening at the rear of the drip tray.
•
Route tubing from ports 2, 8 (B), 3, 4, and 5 up to the left-hand side
of the outlet valve and then down through the opening at the rear of
the drip tray.
•
After routing the tubing, check to ensure that the outer door of the
CM-30S still closes completely.
Route these
tubes up and over
the outlet valve.
Route these tubes up
and to the left-hand side
of the outlet valve