3 Maintaining the 30-cm single-zone column manager
72
October 7, 2014, 715004178 Rev. B
7. Roll a nonabrasive, lint-free wipe, and use it to absorb the liquid from
the leak sensor reservoir and its surrounding area.
8. With a cotton swab, absorb any remaining liquid from the corners of the
leak sensor reservoir and its surrounding area.
9. Align the leak sensor’s T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
10. Reattach the connector to the receptacle on the front of the module, and
power-on the CM-30S.
11. In the instrument console, select Column Manager from the system tree.
Slot in leak
sensor reservoir
T-bar
Leak sensor
connector receptacle