2 Preparing and operating the CM-30S
36
October 7, 2014, 715004178 Rev. B
Warning:
To avoid releasing solvent vapors into the room, route
vent tubing from the waste container to a fume hood or other,
suitable exhaust system.
Warning:
To avoid spills, empty the waste container at regular
intervals.
Notice:
To avoid fluid backup, ensure proper drainage of waste:
• Place the waste container below the system stack.
• Ensure the waste tube does not crimp or bend. A crimp or bend
can impede flow to the waste container.
• Ensure the exit of the waste tube is not immersed in waste
solvent. If necessary, shorten the tube so that no portion of it
drops below the top of the waste container.