Assembling new fittings
October 7, 2014, 715004178 Rev. B
43
Assembling new fittings
For fitting assemblies that use ferrules not previously assembled or set to
tubing, you must mark the compression screw and fitting body, and ensure
the two marks align when you tighten them.
First use
Stainless steel (gold-plated) fitting with
long flats and 2-piece stainless steel
ferrule
Finger-tight, plus 3/4-turn
using wrench
Reinstal
led
Stainless steel (gold-plated) fitting with
long flats and 2-piece stainless steel
ferrule
Finger-tight, plus up to
1/6-turn using wrench
First use or reinstalled
Dual-threaded, finger-tight fitting with
cap nut
1. Loosen the cap nut from
the dual-threaded,
finger-tight fitting.
2. Finger tighten the
dual-threaded fitting with
the ferrule, into the
column.
3. Tighten the cap nut onto
the dual-threaded,
finger-tight fitting.
Installation recommendations for ACQUITY APC and UPC
2
column-compartment fittings: (Continued)
Fitting
Recommended tightening
TP03378
Long flats
2-piece ferrule
3/4-turn
TP02713
Long flats
2-piece ferrule
1/6-turn
Dual-threaded
fitting
Ferrule
Cap nut