Replacing the ferrule on the column fitting
October 7, 2014, 715004178 Rev. B
121
3. Remove the ferrule from the fitting.
4. Install the new ferrule on the fitting.
5. Slide the fitting and ferrule back onto the tubing.
Result:
The ferrule locks onto the fitting.
Fitting
Ferrule
Fitting
Ferrule
Fitting
Tubing