3 Maintaining the 30-cm single-zone column manager
88
October 7, 2014, 715004178 Rev. B
6. Note the number associated with the position of the column-outlet
tubing that you are replacing, and then use the 1/4-inch open-end
wrench to disconnect the column outlet fitting from the outlet valve of
the CM-30S.
Column-outlet valve:
7. Pull the tubing down and through the opening at the rear of the drip
tray.
Warning:
To avoid spills, do not remove the valves' drip tray when
removing the column-outlet tubing. All drip trays must remain in
place during the CM-30S's operation and maintenance.
Column-outlet valve
Opening
Underside of
drip tray