B Specifications
144
October 7, 2014, 715004178 Rev. B
Temperature distribution along
column
+/- 3.0 °C
Test conditions:
• Column not installed.
• No flow.
• Measurement taken relative to
compartment air temperature
sensor.
• Test conducted at 60 °C column
compartment temperature.
• External ambient temperature is
25 °C.
Solvent temperature difference (at
column inlet)
+/-5.0 °C, max.
Test conditions:
• Measurement taken relative to
compartment air temperature
sensor.
• Test conducted at 60 °C column
compartment temperature.
• External ambient temperature is
25 °C.
Column compartment temperature
stability
+/- 0.5 °C
Test conditions:
• Column not installed.
• No flow.
• Measurement taken with
compartment temperature sensor.
• Measurement taken with internal
temperature sensor after 1 hour
thermal equilibrium at set point.
• Test conducted at 60 °C column
compartment temperature.
• External ambient temperature is
25 °C.
Performance specifications: (Continued)
Item
Specification