2 Preparing and operating the CM-30S
42
October 7, 2014, 715004178 Rev. B
When tightening system fittings, consult the following table.
Installation recommendations for ACQUITY APC and UPC
2
column-compartment fittings:
Fitting
Recommended tightening
First use
Stainless steel (gold-plated) fitting with
long flats and 2-piece stainless steel
ferrule on hypo-extended tubing
1. Using your fingers, screw
the gold fitting into the
port and stop as soon as
you begin to feel
resistance (nut is touching
ferrule).
2. Using a 1/4-inch open-end
wrench, tighten until you
begin to feel an increase in
resistance.
3. Using the wrench, tighten
an additional 1/2-turn.
Re-installed
Stainless steel (gold-plated) fitting with
long flats and 2-piece stainless steel
ferrule on hypo-extended tubing
Finger-tight, plus up to
1/6-turn using wrench
TP03493
Long flats
2-piece ferrule
Hypo-extended
tubing
1/2-turn
TP03493
Long flats
2-piece ferrule
Hypo-extended
tubing
1/6-turn