Performance specifications
October 7, 2014, 715004178 Rev. B
143
Time to temperature from off state
45 minutes, max.
Test conditions:
• Column not installed.
• No flow.
• Measurement taken with internal
temperature sensor after 1 hour
thermal equilibrium at set point.
• Test conducted at 60 °C column
compartment temperature.
• External ambient temperature is
25 °C.
Column compartment temperature
accuracy (total error vs. a traceable
standard)
+/- 0.5 °C
Test conditions:
• Column not installed.
• No flow.
• Measurement taken on the
outward-facing side of the
compartment thermistor, in close
contact with the middle of the
length of the thermistor.
Column compartment temperature
precision (repeatability)
+/- 0.1 °C
Test conditions:
• Column not installed.
• No flow.
• Measurement taken at
compartment air temperature
sensor.
Performance specifications: (Continued)
Item
Specification