Replacing the column-outlet tubing in a CM-30S installed in a UPC
2
system
October 7, 2014, 715004178 Rev. B
91
To install the new column-outlet tubing:
1. Thread the valve-fitting end of the tubing through the opening at the
back of the valve’s drip tray.
Tip:
Straighten the tubing slightly, if necessary.
2. Connect the new column-outlet tubing to the appropriate position on the
outlet valve, and tighten the fitting, finger-tight plus an additional
3/4-turn, using the 1/4-inch open-end wrench.
Warning:
To avoid spills, do not remove the valves' drip tray when
removing the column-outlet tubing. All drip trays must remain in
place during the CM-30S's operation and maintenance.
Opening
Underside of
drip tray