Replacing the column-outlet tubing in a CM-30S installed in a UPC
2
system
October 7, 2014, 715004178 Rev. B
89
8. If no column is installed on the column-outlet tubing that you are
replacing, remove the tubing from the column clip.
9. If a column is installed on the column-outlet tubing that you are
replacing, do as follows:
a. Loosen the cap nut from the column fitting, taking care not to
remove the cap nut.
Column-outlet tubing
Column plate
Column clip
Cap nut
Column fitting
Column