Resolving leak sensor errors
October 7, 2014, 715004178 Rev. B
71
4. Detach the leak sensor connector from the front of the CM-30S by
pressing down on its tab and pulling the connector out of the receptacle.
5. Remove the leak sensor from its reservoir, grasping it by its serrations,
and pulling upward.
Leak sensor:
6. Use a nonabrasive, lint-free wipe to dry the leak sensor prism.
Notice:
To avoid damaging the leak sensor, do not grasp it by the
ribbon cable.
Press down on tab to
release connector
Serrations
TP02891
Prism
Lint-free wipe