Replacing the column-outlet tubing in a CM-30S installed in an APC system
October 7, 2014, 715004178 Rev. B
79
4. Grasp the handle of the column-compartment drawer, and pull the
drawer toward you.
5. Use the 1/4-inch open-end wrench to disconnect the column outlet fitting
from the primary column manager outlet valve.
Column-compartment
drawer
Drawer handle