B Specifications
142
October 7, 2014, 715004178 Rev. B
Number of columns
APC
• Maximum of two selectable banks
of columns. Each bank contains a
maximum of four columns.
UPC
2
• Maximum of eight selectable
columns when bypass and waste
lines are omitted.
• Maximum of six selectable columns
with bypass and waste lines in use.
Column compartment settable
temperature range and resolution
All columns mounted in a single
temperature-controlled compartment,
from 20.0 °C to 90.0 °C, in increments
of 0.1 °C
Column compartment controllable
temperature range
From 5 °C above ambient to 90 °C
Time to temperature from steady
state after door is open for 30 s
12 minutes, max.
Test conditions:
• Column not installed.
• No flow.
• Measurement taken with internal
temperature sensor after 1 hour
thermal equilibrium at set point.
• Door open for 30 s.
• Test conducted at 60 °C column
compartment set point
temperature.
• External ambient temperature is
25 °C.
Performance specifications: (Continued)
Item
Specification