Replacing the leak sensor
October 7, 2014, 715004178 Rev. B
75
4. Remove the leak sensor from its reservoir, grasping it by its serrations,
and pulling upward on it.
Leak sensor:
5. Unpack the new leak sensor.
6. Align the leak sensor’s T-bar with the slot in the side of the leak sensor
reservoir, and slide the leak sensor into place.
Notice:
To avoid damaging the leak sensor, do not grasp it by the
ribbon cable.
Serrations
Slot in leak
sensor reservoir
T-bar