128
*PSC {0|1}
Power-on status clear. Clear the Status Byte and Standard Event register enable masks when
power is turned on (*PSC 1). When *PSC 0 is in effect, the Status Byte and Standard Event
register enable masks are not cleared when power is turned on. [Stored in non-volatile
memory]
*PSC?
Query the power-
on status clear setting. Returns “0” (*PSC 0) or “1” (*PSC 1).
*SRE <enable value>
Enable bits in the Status Byte enable register.
*SRE?
Query the Status Byte enable register. The multimeter returns a decimal value which
corresponds to the binary-weighted sum of all bits set in the register.
*STB?
Query the Status Byte summary register. The *STB? command is similar to a serial poll but it
is processed like any other instrument command. The *STB? command returns the same result
as a serial
poll but the “request service” bit (bit 6) is not cleared if a serial poll
has occurred.
B.11 SCPI Compliance Information
This section encloses a list of commands that are device-specific to the DL-1060. Although not
included in the 1999.0 version of the SCPI standard, these commands are designed with the
SCPI format in mind and they follow all of the syntax rules of the standard.
Many of the required SCPI commands are accepted by the multimeter but are not described in
this manual for simplicity or clarity. Most of these non-documented commands duplicate the
functionality of a command already described in this chapter.
MEASure:
CONTinuity?
DIODe?
SAMPle:
COUNt {<
value
>|MINimum|MAXimum}
COUNt? [MINimum|MAXimum]
[SENSe:]
FUNCtion "CONTinuity"
FUNCtion "DIODe"
FREQuency:VOLTage:RANGe {<
range
>|MINimum|MAXimum}
FREQuency:VOLTage:RANGe? [MINimum|MAXimum]
FREQuency:VOLTage:RANGe:AUTO {OFF|ON}