The following conditions will clear the summary register.
⚫
Users execute a *CLS (clear status) command.
⚫
To query the standard event and questionable data registers will clear only the separate
bits in the summary register.
The following conditions will clear the enable register.
⚫
Users turn on the power, and have set the DMM previously by using the *PSC 1 command.
⚫
Users execute a *SRE 0 command.
※
Note: If you have set the DMM previously by using *PSC 0, the status byte enable register
won
’
t be cleared at power-on.
About the Standard Event Register
The standard event register reports the instrument event types below, such as power-on
detected, command syntax errors, command execution errors, self-test (calibration errors),
query errors or the moment of executing an *OPC command. Then through the enable
register, all conditions will be reported in the standard event summary bit. At the moment,
users have to write a decimal value by using the *ESE command to set the enable register
mask.
※
Note 1: The errors in the DMM
’
s error queue except reading the error queue by using
SYSTem:ERRor? will be recorded by an error condition (including bits 2, 3, 4 or 5 of the
standard event register).
※
Note 2: In both the standard event register bit 3 and the questionable data event register
bits 0, 1 or 9, a reading overload condition can be reported all the time. But in the DMM
’
s
error queue, no error message will be recorded.
Bit Definitions for Standard Event Register
Bit
Decimal
Value
Definition
0.
Operation Complete
1
All commands prior to and including an *OPC command have been
executed
1.
Not Used
2
Set to 0.
2.
Query Error
4
The DMM output buffer was empty as the DMM tried to read from
it. Or a new command line was received before a previous query
has been read. Or, both the input and output buffers are full.
3.
Device Error
8
Error occurrence from a self-test, calibration or reading overlaod.
4.
Execution Error
16
Error occurrence from an execution.
5.
Command Error
32
Error occurrence from a command
’
s syntax.
6.
Not Used
64
Set to 0.
7.
Power On
128
Since the last time the event register was read or cleared, power
has been turned off and on.
The following conditions will clear the standard event register.
⚫
Users send a *CLS command.
⚫
Users query the event register by using the *ESR? command.