Error Message
Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the
first error that was stored. When user has read all errors from the queue, the
ERROR
annunciator turns off. DL-1060 beeps once each time an error occurs.
Should more than 20 errors have existed, the last error stored in the queue (the most
recent error) is replaced with -350,
“
Too many errors
”
. No additional errors are saved
until you remove errors from the queue, the multimeter responds with +0,
“
No error
”
.
The error queue is cleared when power has been off or after a *CLS (clear status)
command has been executed. The *RST (reset) command does not clear the error
queue.
5.1 Error Type
There is only one type over here,
“
execution errors
”
.
5.1.1 Execution Errors
⚫
-101 Invalid character
An invalid character was found in the command string.
⚫
-102 Syntax error
Invalid syntax was found in the command string.
⚫
-103 Invalid separator
An invalid separator was found in the command string.
⚫
-104 Data type error
A parameter type error was found in the command string.
⚫
-105 GET not allowed
A Group Execute Trigger (GET) is not allowed in the command string.
⚫
-108 Parameter not allowed
More parameters were found than needed for the command.
⚫
-109 Missing parameter
Not enough parameters were received for the command.