84
check the error message, such as “ERROR CODE” or “NO ERRORS” as shown in
Figure 3-38
.
Figure 3-38
3.4.8 Firmware Revision
DL-1060 has three microprocessors for various internal systems. You can query the
multimeter to determine which revision of firmware is installed for each icroprocessor.
How to check the firmware revision
1.
Do the procedure SHIFT + CONFIG > SYSTEM > SYSTEM VER > ENTER to check
the DL-
1060’s firmware version. The multimeter gives three numbers in xx
-xx-xx
format on the display. The first number is the firmware revision number for the
measurement microprocessor; the second number is for I/O processor and the
third number is for the front-panel processor as shown in
Figure 3-39
.
Figure 3-39
3.4.9 Calibration
※
Notice:
Do not use
“
CALIBRATE
”
since it is aimed for maker calibration.