63
Front Panel Operation
1.
Press SHIFT+CONFIG buttons as shown in
Figure 3-20
.
2.
Set a defined parameter via the procedure TRIG SYS > TRIG CNT > ENTER.
3.
Set the desired number by pressing
△
and
▽
buttons to increase or decrease the
number, and using
◁
and
▷
buttons to move through the digits. Press ENTER to
set the number.
Figure 3-20
Remote Interface Operation
Use the following command to set the number of samples from your PC terminal.
SAMPle:COUNt <value>
B. Number of triggers
Although the meter normally takes one trigger before returning to the “idle” state,
you
can manually specify the number of triggers it acce
pts before the “idle” state.
However, this can only be done through the remote interface. The following command
shows how to set multiple triggers before returning to idle state.
TRIGger:COUNt <value>
C. Reading hold (Front panel operation only)
The reading hold feature is used to hold a stable reading on the display. When a
reading is stable, enabling the reading hold will hold the stable reading and trigger a
beeping sound. Reading hold feature depends on an adjustable sensitivity band to
determine if a reading is stable. The multimeter considers a reading stable when three
consecutive readings are within Sensitivity Band (Hold) will explain how to adjust the
sensitivity band. This feature is only available from the front panel operation.