2.5 Frequency & Period Measurements
DL-1060 uses an on-board counter with 25MHz to measure the frequency / period.
The measurement band is from 3Hz to 300kHz (or 333 ms to 3.3
μs) and the
measurement voltages range from 100mV to 750 V in AC. The range default setting is
auto-range.
The maximum input allowed is 750V. Applying excess voltage may
damage your meter.
How to measure frequency
1.
Connect the test leads to the terminals on the front panel as shown in
Figure 2-6
.
2.
Set RESOLUTION (refer to 3.1.3 Resolutin & NPLC Setting) or skip this step to use
the default setting.
3.
Press FREQ button for frequency measurement.
4.
Press CONFIG button and use
◁
and
▷
buttons to locate the “FREQ” option. Press
ENTER to select Frequency measurement.
5.
Select the auto-range function by pressing AUTO button on the front panel or use
△
and
▽
buttons to manually select the desired range.
6.
Connect test leads to your source signal and observe readings shown on the
display. If the input signal voltage is beyond the allowed range, an overflow
message “
VAC
OVLD” will be displayed.
Figure 2-6
WARNING
!