104
Capacitance Characteristics
Accuracy ± (% of r % of range) [1]
Function
Range
Test
Current
1 Year (23
℃
±
5
℃
)
Capacitance
[6]
1 nF
10
μ
A
2.0+0.80
10 nF
10
μ
A
1.0+0.50
100 nF
100
μ
A
1.0+0.50
1
μ
F
100
μ
A
1.0+0.50
10
μ
F
100
μ
A
1.0+0.50
100
μ
F
1 mA
1.0+0.50
1 mF
1 mA
1.0+0.50
10 mF
1 mA
2.0+0.50
[1] Specifications are for 2-hour warm-up condition, 6 1/2 digits and they
’
re relative to
calibration standards.
[6] The Null function must be used.
Thermocouple Characteristics
[1]
Function
Type
Range
1 Year (23
℃
± 5
℃
)
Thermocouple
[7]
B
600
℃
to 1820
℃
1.5
℃
C
0
℃
to 2316
℃
1.5
℃
E
-250
℃
to 1000
℃
1.5
℃
J
-210
℃
to 1200
℃
1.0
℃
K
-200
℃
to 1372
℃
1.0
℃
N
-200
℃
to 1300
℃
1.0
℃
R
0
℃
to 1767
℃
1.5
℃
S
0
℃
to 1767
℃
1.5
℃
T
-250
℃
to 400
℃
1.5
℃
[1] Specifications are for 2-hour warm-up condition, 6 1/2 digits and they
’
re relative to
calibration standards.
[7] The measurement accuracy excluded the error of test leads.