SENSe:CURRent:DC:NPLCycles
{0.001|0.006|0.02|0.06|0.2|0.6|1|2|10|100|MINimum|MAXimum}
SENSe:CURRent:DC:NPLCycles? [MINimum|MAXimum]
SENSe:RESistance:NPLCycles
{0.001|0.006|0.02|0.06|0.2|0.6|1|2|10|100|MINimum|MAXimum}
SENSe:RESistance:NPLCycles? [MINimum|MAXimum]
SENSe:FRESistance:NPLCycles
{0.001|0.006|0.02|0.06|0.2|0.6|1|2|10|100|MINimum|MAXimum}
SENSe:FRESistance:NPLCycles? [MINimum|MAXimum]
For frequency and period measurements, aperture time (or gate time) is analogous to
integration time, and you can use the following commands to set it. Specify 10 ms (4
1/2 digits), 100 ms (default; 5 1/2 digits), or 1 second (6 1/2 digits).
SENSe:FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
SENSe:FREQuency:APERture? [MINimum|MAXimum]
SENSe:PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}
SENSe:PERiod:APERture? [MINimum|MAXimum]