SENSe:TEMPerature:SPRTD:CX {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:DX {<
value
>|MINimum|MAXimum}
Thermocouple
Definition
DL-1060 is built-in cold junction compensation that can improve the accuracy of thermo
measurements. If you are using this function, you have to set an adapter type. For
example, through the
“
TYPE
”
selections you are available to use a specific adapter K, J, R,
S, T, E, N, B, or C for temperature measurement.
Perhaps it’s not the solution you want according to your application. Pick the next
selection RJUNCTION’s REAL or RJUNCTION’s SIMULATED by the procedure SHIFT > TEMP
> CONFIG > RJUNCTION > REAL/SIMULATED. This procedure is not for the settings, but
the function execution. Before executing, please go setting REAL or SIMULATED
peremeters by the procedure SHIFT > TEMP > CONFIG > SIMULATED/REAL.
What’s the difference between the REAL and the SIMULATED? For the REAL, the REAL
value is a standard value detected from the present internal equipment, i.e. the value
would be frequently influenced and changed at different environments and climates, and it
can’t be adjusted. For the SIMULATED, the SIMULATED value is a reference value. To use
this function, you need to be aware that the temperature value from a certain object is
absolute meaningful and accurate, for example 0
℃
(mixed ice and water). After you well
selet the type and matched the correct adaptor, the tested result from the mixed ice and
water is 2
℃
. At this circumstance, you know the real and tested error is 2
℃
. So the next
step is to adjust the default value from 25
℃
to 23
℃
. Then your following thermocouple
measurements with the simulated function would be accurate.
How to set up thermocouple
You can set up the thermocouple configuration either through the front panel
operation or through the remote interface operation.
Front Panel Operation (Unit and Thermocouple configuration)
1.
Press SHIFT+ TEMP to choose thermocouple function as shown in
Figure 3-12
2.
Press CONFIG, and then use
◁
and
▷
to toggle between options.
3.
To configure unit or thermocouple type, press ENTE
R on “UNIT” and
“
TYPE”
respectively.