103
Frequency and Period Characteristics
Accuracy ± (% of reading) [1]
Function
Range
[
2
]
Frequency (Hz)
1 Year (23
℃
± 5
℃
)
Frequency
&
Period
100 mV
to
750 V[4]
10 to 40
0.03
40 to 300 k
0.02
[1] Specifications are for 2-hour warm-up condition, 6 1/2 digits and they
’
re relative to
calibration standards.
[2] 20
%
over range on all ranges except 750Vac range.
[4] The Range 750Vac is limited to 100kHz.
AC Characteristics
Accuracy ± (% of r % of range) [1]
Function
Range
[
2
]
Resolution
Frequency(Hz) 1 Year (23
℃
± 5
℃
)
ACV
(True RMS
Voltage)
[5]
100.0000 mV
0.1
μ
V
10 to 20 k
0.12 + 0.05
20 k to 50 k
0.25 + 0.05
50 k to 100 k
0.65 + 0.08
100 k to 300 k
4.8 + 0.8
1.000000 V
to
750.000 V[4]
1 uV
~
1 mV
10 to 20 k
0.12 + 0.04
20 k to 50 k
0.25 + 0.05
50 k to 100 k
0.65 + 0.08
100 k to 300 k
4.8 + 0.8
ACI
(True RMS
Current)
[5]
1.000000 A
1
μ
V
10 to 1 k
0.2 + 0.04
1 k to 5 k
1.00 + 0.1
3.00000 A
10
μ
V
10 to 1 k
0.3 + 0.06
1 k to 5 k
1.5 + 0.15
10.00000 A
10
μ
V
10 to 1 k
0.5 + 0.12
1 k to 5 k
2.5 + 0.2
[1] Specifications are for 2-hour warm-up condition, 6 1/2 digits and they
’
re relative to
calibration standards.
[2] 20
%
over range on all ranges except 750Vac range.
[4] The Range 750Vac is limited to 100kHz.
[5] Specifications are for sine wave input>5% of range. (3 Hz Bandwidth)