52
Choosing
USER
takes you to a menu where you can specify factors used in the
calculation equation to obtain the temperature. Use
◁
and
▷
to move through the
digits and
△
and
▽
to change the numbers to a desired value. Press ENTER to set the
value.
Choosing
SPRTD
takes you to a menu where you can specify the seven coefficients
that are used to determine the temperature. Use
◁
and
▷
to move through the
digits and
△
and
▽
to change the numbers to a desired value. Press ENTER to set the
value.
2-wire/4-wire selection
1.
Press TEMP button to select RTD temperature measurement function as shown in
Figure 3-11
.
2.
Press CONFIG button.
3.
Use
◁
and
▷
to locate TRANSDUCER submenu. Press ENTER to select it.
4.
Use
◁
and
▷
to locate a desired option. Press ENTER to finish the process.
Figure 3-11
Remote Interface Operation
Use the following commands to set up the RTD configuration.
SENSe:TEMPerature:RTD:TYPE{PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}
SENSe:UNIT {Cel|Far|K}
SENSe:UNIT?
SENSe:TEMPerature:RTD:RZERo {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:ALPHa {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:BETA {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:DELTa {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:RZERo {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:A4 {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:B4 {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:AX {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:BX {<
value
>|MINimum|MAXimum}