83
3.4.6 Language
DL-1060 supports two languages: DEFAULT (DL-1060) and COMPATIBLE(Keysight
34401A).
How to set up the language
1.
Do the procedure SHIFT + CONFIG > SYSTEM > ENTER > LANGUAGE > ENTER >
DEFAULT/COMPATIBLE > ENTER to select the language as shown in
Figure 3-37
.
Figure 3-37
3.4.7 Error Condition
The error annunciator on the front panel display tells about the error condition of the
multimeter. If there are one or more syntax or hardware errors found, the error
annunciator will be lit to inform you. The multimeter stores errors using the
first-in-first-out (FIFO) order and it records up to 20 errors in the error queue. The
first error detected will be the first error retrieved. In addition, the general conditions
of the ERROR are listed in the following. For more ERROR information, refer to
Chapter 5 for the list of error messages.
※
Note:
⚫
The error annunciator will be turned off when you have read the errors in the
memory.
⚫
If there is no error found when you check the error queue, the multimeter shows
“
NO ERRORS
”.
⚫
If there are more than 20 errors found, the last error in the error queue will be
replaced with “
-350
”
, means too many errors.
⚫
The error queue is cleared when the multimeter is turned off, or when a *CLS
command is received.
How to check the error queue
1.
Do the procedure SHIFT + CONFIG > SYSTEM > ENTER > SCPI ERR > ENTER to