45
Figure 3-6
※
Note: When using the Way One to set the resolution, your options with each press
on the same button are 4 1/2, 5 1/2 and 6 1/2. And under tha DCV, DCI,
Ω
2 &
Ω
4
functions, the NPLC just can be set by pressing CONFIG > NPLC > ENTER >
◁
and
▷
> Target NPLC.
Way Two
(for DCV, DCI,
Ω2
&
Ω4
functions only)
:
1.
Select a desired function within DCV, DCI,
Ω
2 &
Ω
4.
2.
Do the procedure CONFIG > NPLC > ENTER >
◁
and
▷
> Target NPLC to set
NPLC. (Please refer to the
Table 3-1
to set digits.)
Figure 3-7
Remote Interface Operation
Use the following commands on your PC terminal to set the resolution for your
measurement.
CONFigure:<function> <range>,<resolution>
MEASure:<function>? <range>,<resolution>
SENSe:<function>:RESolution <resolution>
You can specify the resolution in the same units as the measurement function, not in
number of digits. For instance, specify the resolution in volts and the frequency in
herts. Below are some examples.
CONF:VOLT:DC 10,0.001 (4 1/2 digits on the10Vdc range)
MEAS:CURR:AC? 1,1E-6 (6 1/2 digits on the1A range)
CONF:FREQ 1kHz,0.1Hz (1000 Hz input, 0.1Hz resolution)
VOLT:AC:RES 0.05 (50mV resolution on the ac function)