66
Front Panel Operation
1.
Press SHIFT+CONFIG buttons as shown in
Figure 3-22
.
2.
Set a difined parameter via the path TRIG SYS > DELAY > MANUAL > ENTER.
3.
Use
◁
and
▷
buttons to move through the digits and
△
and
▽
buttons to
increase or decrease the numbers to set the desired delay time (between 0 to
3600 second). Then press ENTER to actuate the delay time.
Figure 3-22
Remote Interface Operation
You can also use the remote interface operations from your PC terminal to set the
trigger delay. To select a delay time, use the following command or set the trigger
delay to be automatic:
TRIGger:DELay {<seconds>|MINimum|MAXimum}
or
TRIGger:DELay:AUTO {OFF|ON}
3.3 Math Operations
This section will introduce the mathematical operations in DL-1060. There are eight
math operations: PERCENT, AVERAGE, NULL, LIMITS (HIGH LIMIT/LOW LIMIT), MX+B,
dB and dBm testing. They either store data for later use or perform mathematical
operations on the readings. Note that these math operations are available to all
measurement functions except diode and continuity testing.
※
Note: The “MATH” anunnciator on the display indicates the state of a mathematical
feature, excluding Ratio, and will be lit when enabling whichever math feature