
58
3.2 Trigger Operations
The triggering system of the DL-1060 will be explained in this section.
DL-1060 provides a various trigger operations such as trigger mode, trigger source,
and different trigger settings for a specific measurement. The selected setup will be
stored into the volatile memory and it returns to the default mode after power off.
Following
Figure 3-17
describes the detail of the trigger setup.
Figure 3-17
The trigger system is supposed to be in an idle status when DL-1060 is not measuring.
( *(sampling)display will illuminates when measurement is in progress )
When starting measurement after idle status, the trigger system should be in trigger
waiting status. To make is in trigger waiting status, execute one of the following
command via remote interface.
INTiate
READ?
MEASure?
While DL-1060 is in trigger waiting status(Waiting for event), the measurement
operation will starts after execution of trigger delay, when the trigger condition
matches by detecting trigger signal from the selected trigger source etc. The
operation form the front panel will always be trigger waiting status then receives
trigger signal at any time.