64
How to enable/disable reading hold
1.
Press SHIFT+TRIGGER buttons.
2.
To disable it, press SHIFT+TRIGGER buttons again as shown in
Figure 3-21
.
Figure 3-21
D. Trigger delay
This feature is useful for user who needs a longer delay time to wait for the system to
be stabilized. The time needed to stabilize a measurement system is called the
“settling time.” The settling time depends on the measurement range, cable
properties and signal source.
Defaults
The default of the trigger delay is automatic. DL-1060 automatically selects a delay
time for you according to the setting of the measurement if you do not specify a delay.
A list of the default for each measurement function is shown on
Table 3-4
. The range
for the delay is from 0 to 3600 seconds. The defined selection of delay time is stored
in a volatile memory and the default will be restored when the meter is power-off.
The internal settings of automatic trigger delay
The delay time for automatic trigger delay is determined by the selection of
measurement function, range, and the setting of integration time and AC filter speed.
The following table shows the default delay time for each measurement setting.