Appendix B: Performance Verification
BĆ26
DG2040 Data Generator
This test verifies that the internal trigger generator and the external trigger input
are functioning correctly.
The following equipment is required.
Equipment Required
Oscilloscope
Function generator
Two BNC cables
Performance check disk
Connections.
Refer to Figure B-14 for connections.
Use the BNC cables to make the following connections:
H
Connect the sync output from the DG2040 Data Generator rear panel to the
oscilloscope CH4 input.
H
Connect the function generator output to the DG2040 Data Generator front
panel trigger input.
TDS784D
DG2040 Data Generator
AFG320
BNC Cable: connect to the
AUX OUTPUT: SYNC
connector on the DG2040 Data Generator rear panel.
CH4
BNC Cable
Trigger Input
CH1
Figure BĆ14: Internal trigger generator & external trigger input connection
Internal Trigger Generator
& External Trigger Input
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com