![Tektronix 59021-10 User Manual Download Page 153](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414153.webp)
Reference
3Ć70
DG2040 Data Generator
Output Condition Menu
The Output Condition menu defines which data bits are assigned to which output
channel. The menu lists the data bits at the right and the output channels at the
left. Refer to Figure 3-58. Use the general-purpose knob to select data bits and
the up and down arrow buttons to select output channels.
The Assign Condition function assigns the selected data bit to the selected
output channel. The currently set value is overwritten.
Output Channel List
Data Bits List
Figure 3Ć58: Output channel assign popĆup menu
See Figure 2-6 for the output channel allocation.
Sub Menu
Item
Function
Assign
Assigns the selected data bit to the selected output channel using the
popĆup menu brought up at the same time as the popĆup menu. Refer
to Figure 3Ć58. The currently set value is overwritten.
Release
Clears the data for the selected assignment.
Operation.
Do the following to assign a data bit to the output channel.
Bottom button
PopĆup menu
Side button
Output Condition
Assign Condition
Select a data bit with the generalĆpurpose knob.
Assign Condition
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com