Operating Basics
2Ć38
DG2040 Data Generator
2. Set the length of pattern memory (the number of samples) to 1440 points.
Menubutton
Bottom button
PopĆup menu
Side button
FrontĆpanel
button
Settings
Set memory
size
OK
1440
ENTER
OK
3. Set the block cursor position and width. This procedure sets a scope of 2
(DATA0 and DATA1) and a width of 360 samples.
Menubutton
Bottom button
PopĆup menu
Side button
FrontĆpanel
button
Execute
Action
Set scope
OK
2
OK
Move the knob icon to the
Width
window in the upper left area of the screen using the
frontĆpanel
CURSOR
button.
360
ENTER
4. Divide the data between block 1 (point 0 to 359) and block 2 (point 360 to
719). Use BK2 as the name for block 2.
Menubutton
Bottom button
PopĆup menu
Side button
FrontĆpanel
button
Move the knob icon to the
Cursor
window in the upper left area of the screen by
pressing the frontĆpanel
CURSOR
button once or twice.
360
ENTER
Block
Add block
delimiter here
OK
Clear String
BK2
OK