![Tektronix 59021-10 User Manual Download Page 240](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414240.webp)
Appendix B: Performance Verification
DG2040 Data Generator
BĆ47
Table BĆ9: DG2040 test record (cont.)
Rise time and fall time
accuracy
High limit
Test result
Low limit
Frequency
CLOCK sig al
100 MHz
< 150 ps
CLOCK signal
Fall time
500 MHz
< 150 ps
Fall time
1100 MHz
< 150 ps
CH0 sig al
100 MHz
< 150 ps
CH0 signal
Rise time
500 MHz
< 150 ps
Rise time
1100 MHz
< 150 ps
CH0 sig al
100 MHz
< 150 ps
CH0 signal
Fall time
500 MHz
< 150 ps
Fall time
1100 MHz
< 150 ps
CH0 sig al
100 MHz
< 150 ps
CH0 signal
Rise time
500 MHz
< 150 ps
Rise time
1100 MHz
< 150 ps
CH0 sig al
100 MHz
< 150 ps
CH0 signal
Fall time
500 MHz
< 150 ps
Fall time
1100 MHz
< 150 ps
CH1 sig al
100 MHz
< 150 ps
CH1 signal
Rise time
500 MHz
< 150 ps
Rise time
1100 MHz
< 150 ps
CH1 sig al
100 MHz
< 150 ps
CH1 signal
Fall time
500 MHz
< 150 ps
Fall time
1100 MHz
< 150 ps
CH1 sig al
100 MHz
< 150 ps
CH1 signal
Rise time
500 MHz
< 150 ps
Rise time
1100 MHz
< 150 ps
CH1 sig al
100 MHz
< 150 ps
CH1 signal
Fall time
500 MHz
< 150 ps
Fall time
1100 MHz
< 150 ps
1
Ta = Degree C = (Fahr 32)5/9
2
Low limit = |Ta - 25 degree C| X 15 ps - 100ps
3
High Limit = |Ta - 25 degree C| X 15 ps + 100ps
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com