![Tektronix 59021-10 User Manual Download Page 198](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414198.webp)
Appendix B: Performance Verification
DG2040 Data Generator
BĆ5
Self Tests
Execute the DG2040 Data Generator self tests and verifies that no errors
occurred.
1. Press the required buttons in the following order. Selecting All from the
Type side button causes the instrument to run the self-test on all test items.
Menubutton
Bottom button
Pop up menu
Side button
Front panel
button
UTILITY
Diag
Type (Select All)
Execute
2. Verify that all self tests passed by checking the Status display area shown in
Figure B-2.
Figure BĆ2: Diagnostics menu
If Fail is listed in the Status display area, an error code will be displayed in
the Comment column. See Table B-3 for the meanings of the error codes.
NOTE
. Contact your Tektronix sales representative if an error occurs.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com