![Tektronix 59021-10 User Manual Download Page 164](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414164.webp)
Reference
DG2040 Data Generator
3Ć81
1 V
0 V
-1 V
Input signal
Transfer function
Time
The highlighted edge controlled
item moves according to the
input signal voltage
Time
Figure 3Ć66: Relation between incoming signal and edge position shift
You can also set the edge offset position within the range from –100 ps to 100 ps
with the Position Offset side menu. Both the incoming signal voltage and the
setting in the Position Offset side menu affect the specified edge to be moved.
Figure 3-67 shows an example of the edge control window.