![Tektronix 59021-10 User Manual Download Page 205](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414205.webp)
Appendix B: Performance Verification
BĆ12
DG2040 Data Generator
Setup.
Set the oscilloscope and AFG320 Function Generator as indicated below:
H
Oscilloscope
Displayed channel
CH1, CH2, CH3 and CH4
Vertical axis
1 V/div for CH1 and CH2
. . . . .
5 V/div for CH3 and CH4
Horizontal axis
5
m
s/div
. . .
Record length
500
. . . .
Acquire mode
Peak Detect
. . . .
Acquire sequence
RUN/STOP button only
.
Trigger mode
Auto
. . . .
Trigger level
2 V
. . . . .
Trigger source
CH3
. . . .
Trigger position
50 %
. .
Trigger slope
Falling Edge
. . . . .
Input coupling
DC
. . . .
Input impedance
50
W
for CH1 and CH2
. .
1 M
W
for CH3 and CH4
H
Function generator
Output channel
CH1
. . .
Waveform
Square wave
. . . . . . .
Frequency
500 Hz
. . . . . . .
Amplitude
1 V (50
W
termination))
. . . . . . .
Offset
500 mV
. . . . . . . . . .
Procedure.
Follow the steps below to verify pattern data output:
1. Load the TP2PG.PDA test pattern file from the performance check disk.
2. Press the START/STOP button on the front panel. The LED, corresponding
to the button, should be on.
3. Verify that the data pattern shown in Figure B-5 appears on the oscilloscope.
4. Disconnect the SMA cable from the CH0 output connector and the CH0
output connector on the DG2040 Data Generator , and then connect them to
the CH1 output connector and the CH1 output connector on the DG2040
Data Generator.
5. Verify that the pattern data shown in Figure B-5 appears on the oscilloscope.
6. Disconnect the BNC cable from event input connector on the DG2040 Data
Generator rear panel.
7. Change the oscilloscope acquisition sequence to Single Acquisition
Sequence.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com