Appendix B: Performance Verification
DG2040 Data Generator
BĆ3
Table B-2 lists the equipment required for the performance tests.
Table BĆ2: Required equipment
Item
No.
Required precision
Recommended equipment
Frequency counter
1
Frequency range: 0.1 Hz to 1250 MHz
Precision: 7 digits or higher
ANRITSU MF 1603A
Digital multimeter
1
DC voltage range:
±
10 V
Precision: 0.01 V
Tektronix DM2510
Performance check disk
1
Tektronix part no. 063Ć3121ĆXX
(provided with the DG2040 Data Generator)
Oscilloscope
1
Bandwidth: 1 GHz or higher
Tektronix TDS784D
Sampling oscilloscope with
head
1
Bandwidth: 6 GHz or higher
Tektronix 11801C with head SDĆ22
Function generator
1
Amplitude: 4 V, offset: 2 V
(50
W
termination),
Frequency: 1 MHz or higher
Tektronix AFG320
BNC cable
5
Impedance: 50
W
Length: 24 inches
Tektronix part no. 012Ć1342Ć00
SMA cable
5
Impedance: 50
W
Length: 20 inches
Tektronix part no. 174Ć1427Ć00
SMA female-to-BNC male
adapter
2
Tektronix part no. 015Ć0572Ć00
Precision 50
W
termination
1
Impedance: 50
W (0.1 %)
Tektronix part no. 011Ć0129Ć00
SMA X5 attenuatorĂ(14ĂdB)
3
Tektronix part no. 015Ć1002Ć00
NĆtoĆBNC adapter
1
Tektronix part no. 103Ć0045Ć00
BNCĆdualĆbanana adapter
1
Tektronix part no. 103Ć0090Ć00
50
W
SMA termination
2
Tektronix part no. 015Ć1022Ć00
Required Equipment
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com