DG2040 Data Generator
AĆ1
Appendix A: Specifications
General Characteristics
This section describes the general characteristics of the DG2040 Data Generator.
All specifications are guaranteed unless labeled “typical”. Typical specifications
are provided for your convenience but are not guaranteed. Specifications marked
with the
√
symbol are checked in the performance verification procedure
beginning on page B-1.
The certification and compliances for the DG2040 Data Generator are found at
the end of this appendix.
The electrical characteristics are valid under the following conditions:
H
The instrument must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications.
H
The instrument must have had a warm-up period of at least 20 minutes.
H
The instrument must have been calibrated and adjusted at an ambient
temperature b20
_
C and +30
_
C.
H
The instrument must be operating at an ambient temperature between
+10
_
C to +40
_
C, unless otherwise noted.
Table AĆ1: Electrical characteristics
Characteristics
Description
Operation mode
1
Repeat
Pattern data is repeatedly output. When a sequence is specified, patterns are repeated
according to the sequence order. The extended sequence functions such as trigger wait, event
jump, and so on are ignored in this mode.
Single
Pattern data is output only once. When a sequence is specified, a trigger signal outputs
according to the sequence order.
Step
Pattern data is output based on the clock, not specified by the clock source, but generated by
pressing the
STEP/EVENT
button on the front panel. This mode is the same as the Repeat
mode except for the clock.
Enhanced
Pattern data is output completely according to a sequence. All extended sequence functions,
such as trigger wait, and event jump, are valid in this mode. This mode is the same as the
Repeat mode except for the extended sequence functions.
Performance Conditions
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com