Appendix B: Performance Verification
BĆ42
DG2040 Data Generator
11801C
CH2
SMA 5X attenuator
CH1
DG2040 Data Generator
SMA 5X attenuator
TRIGGER
INPUTS
DIRECT
Figure BĆ20: Rise time and fall time measurement connection
Setup.
Set the sampling oscilloscope as indicated below:
H
Sampling oscilloscope
Displayed channel
CH1 and CH2
Vertical axis
50 mV/div for CH1 and CH2
. . . . .
Horizontal axis
200 ps/div
. . .
Record length
500
. . . .
Acquire mode
Average 32
. . . .
Timebase position
0s
.
Vertical offset
100 mV
. . . .
Trigger mode
Norm
. . . .
Trigger level
100 mV
. . . . .
Trigger source
DIRECT
. . . .
Trigger slope
Rising Edge
. . . . .
Procedure.
Follow the steps below to verify the rise and fall time accuracy:
1. Load the TP11RISE.PDA test pattern file from the performance check disk.
2. Press the following buttons to set the DG2040 Data Generator clock
frequency.
Menubutton
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Clock
Int FREQ
100
MHz/
m
s