Appendix B: Performance Verification
DG2040 Data Generator
BĆ29
This test verifies the edge control mode operation is functioning correctly.
The following equipment is required.
Equipment Required
Sampling oscilloscope with head
Function generator
BNC cable
Four SMA cables
SMA femaleĆtoĆBNC male adapter
50
W
termination
Three SMA X5 attenuators
Performance check disk
Connections.
Refer to Figure B-15 for connections.
Use the SMA cables and the SMA 5X attenuators to make the following
connections:
H
Connect CH0 output from the DG2040 Data Generator front panel to the
sampling oscilloscope CH1 input.
H
Connect CH1 output to the DG2040 Data Generator front panel to the
sampling oscilloscope CH2 input.
H
Connect the CH1 output from the DG2040 Data Generator front panel to the
sampling oscilloscope TRIGGER INPUTS DIRECT.
H
Use the remaining SMA cable, the SMA female-to-BNC male adapter, and
the 50
W
termination to terminate the DG2040 Data Generator CH0 output.
Use the BNC cable to make the following connection:
H
Connect the function generator CH1 output to the DG2040 Data Generator
front panel edge control input.
Edge Control Mode
Operation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com